Comparison of central corneal thickness measured by anterior segment optical coherence tomography, pentacam and ultrasonic pachymetry

ADD TO MY LIST
 
Author(s): Wei Zhao | Ting Wu | Ze-Hong Dong | Jie Feng | Yu-Feng Ren | Yu-Sheng Wang

Journal: Guoji Yanke Zazhi
ISSN 1672-5123

Volume: 13;
Issue: 12;
Start page: 2551;
Date: 2013;
VIEW PDF   PDF DOWNLOAD PDF   Download PDF Original page

Keywords: anterior segment optical coherence tomography | pentacam | ultrasonic pachymetry | central corneal thickness | measure

ABSTRACT
AIM: To investigate and analyze the differences in central corneal thickness(CCT)measurement by anterior segment optical coherence tomography(AS-OCT), pentacam rotating Scheimpflug camera and ultrasonic pachymetry(USP). METHODS: The retrospective study measured CCT of 43 patients(86 eyes)performed excimer laser myopia corrective surgery by the same examiner using RTVue AS-OCT with an anterior segment adaptor, pentacam rotating Scheimpflug camera, and PACSCAN300P USP. Data of three groups were compared and statistically analyzed by pair t test.RESULTS: The mean CCT(±SD)measured by AS-OCT,pentacam and USP were 535±52, 550±52, 550±51μm, respectively. Paired-sample t test showed that the results of AS-OCT, pentacam and USP were statistically different. There were no statistically significant differences in the mean CCT between the pentacam and USP group. Significant correlations of measurement result were found between pentacam and USP(r=0.998, P