Uncertainty of atomic force microscopy measurements

ADD TO MY LIST
 
Author(s): Teodor P. Gotszalk | Pawel Janus | Andrzej Marendziak | Roman F. Szeloch

Journal: Optica Applicata
ISSN 0078-5466

Volume: 37;
Issue: 4;
Start page: 397;
Date: 2007;
VIEW PDF   PDF DOWNLOAD PDF   Download PDF Original page

Keywords: uncertainty of measurement | atomic force microscopy (AFM)

ABSTRACT
We consider the problem of uncertainty in geometrically linear measurements in scanning probe microscopy (SPM) represented by atomic force microscopy (AFM). The uncertainties under consideration are associated both with quantum phenomena in the space cantilever tip–sample surfaces and with effects of dynamic behavior of electronic and optic measurement and control systems. In our experiment, we have analyzed uncertainty of calibrated atomic force microscopy (C-AFM) measurement in two dimensions. Uncertainty of measurements has been estimated according to GUM procedure.