Wide-Beam X-Ray Source Target Thermal Management Simulation Using Inner Jet Cooling

ADD TO MY LIST
 
Author(s): Chang H. Kim | J. Michael Doster | Mohamed A. Bourham

Journal: Scholarly Research Exchange
ISSN 1687-8299

Volume: 2009;
Date: 2009;
Original page

ABSTRACT
A wide-beam area X-ray source has been proposed as a practical replacement for synchrotron sources in clinical DEI applications. Due to a wide X-ray illumination area, a decrease in X-ray flux is expected and thus high electron beam currents up to 3A are considered. To ensure the target performance without deterioration, melting, cracking, or even evaporation, an active cooling system is required for the target block in order to remove the heat and allow for sufficient scanning time. In this study, jet cooling of the target back is investigated for a prototype proof-of-principle target. The prototype target was simulated with the transient k-ɛ turbulence multiphysics model in ANSYS CFX. The simulations were conducted at a heat flux of 1.8×107 W/m2, consistent with values anticipated for a full scale target. The simulation results show that the target temperature exceeds the copper melting point in 2 seconds at inlet velocities below 2 m/s. Also, critical heat flux calculations show that a 1.5 m/s inlet velocity at atmospheric pressure is a lower limit for prevention of target burnout using water as a coolant. Inlet velocities in excess of 2 m/s allows for steady-state operation while satisfying all thermal design constraints.